NEXT GENERATION GAS STERILIZATION


 

EOGas AN333 Features and Specifications

 

EOGas Features

  • Our multiload EOGas system adds and removes bags at any time while other bags are processing.
  • Gas diffusion eliminates "dead space" around loads by delivering Ethylene Oxide directly to the load.
  • No harsh preconditioning is required, treating sensitive products gently and effectively. Conventional systems fill the entire chamber with EO, possibly overdosing product to compensate for "dead space"around loads. Forced pre-conditioning in a deep vacuum under high humidity may damage sensitive products.
  • Available abator

Ordering and information:

Call: 1-800-523-1276
Or visit our Contact Page

 

 

Specifications

  Description High-capacity free-standing
sterilizer and extractor unit
  Capacity 33 cu ft
  Cabinet Weight 630 lbs (286 kg)
  Extractor Weight 10 lbs
(4.5 kg)
  Exhaust Requirements Dedicated 3" (7.6 cm)
discharge vent
  Power Requirements 220/240 V, 50/60 Hz, 5.4 Amps




For more information about EOGas Series 3+ , please click here.

 

MADE WITH PRIDE IN THE U.S.A.

ANDERSEN PRODUCTS, INC . :: . TOLL FREE: 800-523-1276 . :: .ONTCONTACT US