Technical
Bulletin
4/04/2007
Subject: LOW TEMP WARNING error screen seen on Anprolene Sterilizers
equipped with operating versions AICUL320 and AICUL321.
Affected Equipment:
AN74i 115V model 74.64 Lot/Serial # from 261181 to 261577.
AN74j 115V model 74.74 Lot/Serial # from 261215 to 261637.
AN74ix 115V model 74.84 Lot/Serial # from 261192 to 261435.
AN74ix 230V model 74.85 Lot/Serial # from 261377 to 261932.
For the above equipment, if a LOW TEMPERATURE condition is
experienced during a sterilization cycle, after removing the
sterilized items, if the sterilizer door is left OPEN when the
exit button is pressed, the following error display will be seen:
VENT PUMP
FAILURE
LOW
TEMP WARNING
SECTION
OF MANUAL
To remove this message, move the power switch to the off position,
then press and hold both function keys while returning the power
switch to the on position. Continue to hold both function keys
until the display reappears, then release the function keys.
The start-up screen should now be seen.
This message will appear ONLY under the above conditions and
will not appear if the sterilizer door is CLOSED before the exit
button is pressed.
Please review the LOW TEMPERATURE
section in the owner’s
manual for information relating to sterilization cycles that
have experienced a low temperature condition.
If you need any additional information please call 1-800-523-1276 for assistance.
|